Description
Calibrator for Thermocouple & RTD, ± (0.008 % of Rdg + 0.006 mV), ± (0.015 % of Rdg + 0.05 Ω), Supports K, J, E, T, R, S, B, N, G, C, D, L, P Type, 10 Ω to 1000 Ω, 50 hours, CL543B Series, Each
Specifications
Manufacturer
Omega Engineering
Component
Handheld Calibrator
Features
Calibrate Thermocouple Instruments to 0.1 and 0.01, °F and °C
Types J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), (T-DIN) and P (Platinel): -13.000 to 80.000 mV
Calibrate RTD Instruments to 0.1 and 0.01, °F and °C
Pt 10, 50, 100, 200, 500 and 1000 (3850)
Additional Features 1Read and Source Signals
AlertsRed LED displays OVERLOAD or out of range conditions
Battery life50 hr, low battery warning with less than 1 hr operation left
Cold Junction Sensor≤± 0.01% °/ ° from ambient temperature ( °C or °F)
Display TypeLCD
Display Type NoteHigh contrast graphic Liquid Crystal Display. LED backlighting for use in low lit areas.
Height41 mm
Length143 mm
Operating Temperature, Min-20 °C
Over-Voltage ProtectionUp to 60 vrms (rated for 30 seconds), protection against misconnection
Relative Humidity, Max0.9
Relative Humidity, Min10%
Relative Humidity Note10% ≤RH ≤90% (0 to 35°C),non-condensing
10% ≤RH ≤70% (35 to 60°C),non-condensing
RTD and OHMS ReadResistance Ranges:
400Ω Range: 0.00 to 401.00 and 0.000 to 401.000
4000Ω Range: 0.0 to 4010.0 and 0.00 to 4010.00
Accuracy: ± (0.015% of reading + 0.05Ω)
Excitation Current: 0.9 mA to 401Ω, 0.4 mA to 4010Ω (nominal)
RTD and OHMS Source3- and 4-Wire Accuracy:
From 1 to 10.2 mA External Excitation Current: ± (0.015% of setting + 0.05Ω)
Below 1 mA of External Excitation Current: Add (0.025 mV/mA excitation current)
2-Wire Accuracy: Add 0.1Ω to 3-wire and 4-wire accuracy
Resistance Ranges:
400Ω Range: 0.00 to 401.00 and 0.000 to 401.000
4000Ω Range: 0.0 to 4010.0 and 0.00 to 4010.00
RMS Noise:
400Ω Range: ≤± 0.005Ω from 0.1 to 10 Hz
4000Ω Range: ≤± 0.05Ω from 0.1 to 10 Hz
Allowable Excitation Current Range:
400Ω Range: 10.2 mA max
steady or pulsed/intermittent
4000Ω Range: 1 mA max
steady or pulsed/intermittent
Pulsed Excitation Current Compatibility: DC to 0.01 second pulse width, 0.025 mV, mA excitation current
Storage Temperature, Max60 °C
Storage Temperature, Min-30 °C
Supply Voltage6 Vdc
Temperature Effect≤± 25 ppm/°C
Thermocouple TypeJ, K, T, E, R, S, B, C, N, L, G, U, D
Voltage ReadRanges and Resolution: -13.000 to 80.000 mV and -13.0000 to 80.0000 mV
Accuracy: ± (0.008% of reading + 0.006 mV)
Input resistance: ≥10 MΩ
Normal Mode Rejection: 50/60 Hz,50 dB
Common Mode Rejection: 50/60 Hz,120 dB
Voltage SourceRanges and Resolution: -13.000 to 80.000 mV and -13.0000 to 80.0000 mV
Accuracy: ± (0.008% of setting + 0.006 mV)
Source Current: ≥10 mA
Output Impedance: <0.3Ω
RMS Noise: ≤± 0.0005 mV from 0.1 to 10 Hz
Short Circuit Duration: Infinite
Weight0.34 kg
Weight Note0.34 kg (12.1 oz) with boot and batteries